Home > Components > Flow Control System

    LSP01-1A/2A Single Channel Syringe Pump

    LSP01-1A & LSP01-2A  laboratorial syringe pumps are single channel syringe pump which only has infusion mode. The acceptable syringe specification is from 10 μL to 60 mL. Suitable f

    1. description

    LSP01-1A & LSP01-2A  laboratorial syringe pumps are single channel syringe pump which only has infusion mode. The acceptable syringe specification is from 10 μL to 60 mL. Suitable for high accuracy and small flow rate liquid transferring.

    Flow rates:0.831nl/min-54.155ml/min

    Max. Number of Channels:1

    Applications:

    For high precision fluid dosing, especially for laboratorial application.



    Product Instruction:

    LSP01-1A & LSP01-2A  laboratorial syringe pumps are single channel syringe pump which only has infusion mode. The acceptable syringe specification is from 10 μL to 60 mL. Suitable for high accuracy and small flow rate liquid transferring.

    Main Functions and Features

    - Syringe Selection: The syringe can be selected in the manufacturer table which includes manufacturer, material and size or input the inner diameter of the syringe barrel directly 
    - Easy to operate: Combining big screen LCD display with rotary coded switch and membrane keypad makes the operation simple and prompt. 
    - Working Mode: Infusion 
    - Memory Function: 
    1. The parameters are saved in EEPROM. The parameters don't need to be reset when power returns after an interruption 
    2. In flow rate mode, the pump remains running or stop according to the setting parameters when power returns after an interruption 
    - Protection Function: The pump will stall and give an alarm when the drive structure of the pump is blocked 
    - Communication Function: Realize computer control through RS485 communication interface 
    - External Control Function: Input/output control 
    - Calibration Function: Acquire accurate volume through calibration 
    - Syringe Protection: Adjust syringe rest to prevent syringe from damaging

    Technical Specifications

    Syringe pumpLSP01-1ALSP01-2A
    No. of Syringes11
    Advance per Mircostep1/16step: 0.156 μm1/16step: 0.031 μm
    Infusion Volumn per Microstep0.13 μL (60 mL BD Syringe)0.026 μL (60 mL BD Syringe)
    Syringe Size10 μL - 60 mL10 μL - 60 mL
    Max. Linear Rate65 mm/min13 mm/min
    Min. Linear Rate5μm/min1μm/min
    Flow Rate0.831 nL/min -  54.155 ml/min0.166 nL/min -  10.83 ml/min
    Working ModeInfusion
    Linear Force> 9 kgf
    Max. Step Rate6933  (1/16step)/ sec
    Min. Step Rate16  (1/16 step) /30 sec
    Accuracy≤ 0.5% error in the condition of > 30% of max. infusion distance
    Setting ModeRotary coded switch and membrane keypad
    Display128 x 64 graphic LCD
    PowerAC 100 - 240 V
    Operating ConditionTemperature  5 °C - 40 °C       Relative humidity < 80 %
    Dimensions280 × 210 × 140 mm280 × 210 × 140 mm
    Weights3.6 kg3.6kg

    Other parameters

    TypeProduct CodeSyringeSyringe ID (mm)Flow Rate (μl/min-ml/min)
    LSP01-1A050340110μl0.500.001-0.0128
    25μl0.800.0025-0.0327
    50μl1.100.0048-0.0618
    100μl1.600.0101-0.1307
    250μl2.300.0208-0.2701
    500μl3.250.0415-0.5392
    1ml4.720.0875-1.1373
    2ml9.000.3181-4.1351
    5ml13.100.6739-8.7608
    10ml16.601.0821-14.068
    20ml19.001.4176-18.429
    30ml23.002.0774-27.006
    60ml29.143.3346-43.349