LSP02-1B Dual Channels Syringe Pump
LSP02-1B is an infusion/withdrawal syringe pump with two channels. The acceptable syringe specification is from 10 ?L to 140 mL. Suitable for high accuracy and small flow rate liquid transferring.Flow
LSP02-1B is an infusion/withdrawal syringe pump with two channels. The acceptable syringe specification is from 10 ?L to 140 mL. Suitable for high accuracy and small flow rate liquid transferring.
Flow rates:0.831nl/min-150.5ml/min
Max. Number of Channels:2
Applications:
For high precision fluid dosing, especially for laboratorial application.
Product Instruction:
LSP02-1B is an infusion/withdrawal syringe pump with two channels. The acceptable syringe specification is from 10 ?L to 140 mL. Suitable for high accuracy and small flow rate liquid transferring.
Main Functions and Features
- Syringe Selection: The syringe can be selected in the manufacturer table which includes manufacturer, material and size or input the inner diameter of the syringe barrel directly
- Easy to operate: Combining big screen LCD display with rotary coded switch and membrane keypad makes the operation simple and prompt.
- Working Mode: Infusion, withdrawal, infusion/withdrawal, withdrawal/infusion, continuous
- Memory Function: 1. The parameters are saved in EEPROM. The parameters don't need to be reset when power returns after an interruption. 2. In flow rate mode, the pump remains running or stop according to the setting parameters when power returns
- Protection Function: The pump will stall and give an alarm when the drive structure of the pump is blocked
- Communication Function: Realize computer control through RS485 communication interface
- External Control Function: Input/output control
- Calibration Function: Acquire accurate volume through calibration
- Syringe Protection: Adjust syringe rest to prevent syringe from damaging
- Easy to operate: Combining big screen LCD display with rotary coded switch and membrane keypad makes the operation simple and prompt.
- Working Mode: Infusion, withdrawal, infusion/withdrawal, withdrawal/infusion, continuous
- Memory Function: 1. The parameters are saved in EEPROM. The parameters don't need to be reset when power returns after an interruption. 2. In flow rate mode, the pump remains running or stop according to the setting parameters when power returns
- Protection Function: The pump will stall and give an alarm when the drive structure of the pump is blocked
- Communication Function: Realize computer control through RS485 communication interface
- External Control Function: Input/output control
- Calibration Function: Acquire accurate volume through calibration
- Syringe Protection: Adjust syringe rest to prevent syringe from damaging
Technical Specifications
Syringe Pump | LSP02-1B |
Number of Channels | 2 |
Display Setting | Display volume, flow rate, linear speed |
Syringe Size | 10?L – 140 mL |
Power-off Memory | Storing the running parameters automatically |
Dimensions (mm) | 280×250×140 |
Weights (Kg) | 4.3 |
Working Mode | Infusion / Withdraw |
Linear Force | > 18Kg |
Advance per Micro-step | 1/16 Step (0.156?m) |
Max. Step Rate | 13867 (1/16 step) /sec |
Min. Step Rate | 16 (1/16) / 30 sec |
Max. Linear Rate | 130 mm/min |
Min. Linear Rate | 5?m/min |
Accuracy | ≤ 0.5% error in the condition of ≥ 30% of max. infusion distance |
Setting Mode | Rotary coded switch and membrane keypad |
Display | 128x64 graphic LCD |
Power Supply | AC100 -240V |
Operation Condition | Temperature: 5℃ - 40℃, Relative humidity: <80% |
Other parameters
Type | Product Code | Syringe | Syringe ID (mm) | Flow Rate (μl/min-ml/min) |
LSP02-1B | 0503441 | 10μl | 0.50 | 0.001-0.0255 |
25μl | 0.80 | 0.0025-0.0653 | ||
50μl | 1.10 | 0.0048-0.1235 | ||
100μl | 1.60 | 0.0101-0.2614 | ||
250μl | 2.30 | 0.0208-0.5401 | ||
500μl | 3.25 | 0.0415-1.0784 | ||
1ml | 4.72 | 0.0875-2.2747 | ||
2ml | 9.00 | 0.3181-8.2702 | ||
5ml | 13.10 | 0.6739-17.522 | ||
10ml | 16.60 | 1.0821-28.135 | ||
20ml | 19.00 | 1.4176-36.859 | ||
30ml | 23.00 | 2.0774-54.012 | ||
60ml | 29.14 | 3.3346-86.699 |