Home > Components > Flow Control System

    LSP10-1B - 10 Channels Syringe Pump

    LSP10-1B is an infusion/withdrawal syringe pump, which could be mounted with 10 syringes at most. The acceptable syringe specification is from 10 ?L to 10 mL. Suitable for high accuracy and small flow

    1. description

    LSP10-1B is an infusion/withdrawal syringe pump, which could be mounted with 10 syringes at most. The acceptable syringe specification is from 10 ?L to 10 mL. Suitable for high accuracy and small flow rate liquid transferring.

    Flow rates:0.831nl/min-21.675ml/min

    Max. Number of Channels:10

    Applications:

    For high precision fluid dosing, especially for laboratorial application.



    Product Instruction:

    LSP10-1B is an infusion/withdrawal syringe pump, which could be mounted with 10 syringes at most. The acceptable syringe specification is from 10 ?L to 10 mL. Suitable for high accuracy and small flow rate liquid transferring.

    Main Functions and Features

    - Syringe Selection: The syringe can be selected in the manufacturer table which includes manufacturer, material and size or input the inner diameter of the syringe barrel directly 
    - Easy to operate: Combining big screen LCD display with rotary coded switch and membrane keypad makes the operation simple and prompt. 
    - Working Mode: Infusion, withdrawal, infusion/withdrawal, withdrawal/infusion, continuous 
    - Memory Function: 1. The parameters are saved in EEPROM. The parameters don't need to be reset when power returns after an interruption. 2. In flow rate mode, the pump remains running or stop according to the setting parameters when power returns 
    - Protection Function: The pump will stall and give an alarm when the drive structure of the pump is blocked 
    - Communication Function: Realize computer control through RS485 communication interface 
    - External Control Function: Input/output control 
    - Calibration Function: Acquire accurate volume through calibration 
    - Syringe Protection: Adjust syringe rest to prevent syringe from damaging

    Technical Specifications

    Syringe PumpLSP10-1B
    Number of Channels10
    Injected Volume per Micro-step0.0257?L (10 mL BD Syringe)
    Syringe Size10?L – 10 mL
    Flow Rate (60 mL Syringe)0.831nL/min – 21.675 mL/min
    Dimensions (mm)280×250×140
    Weights (Kg)4.3
    Working ModeInfusion / Withdraw
    Linear Force> 18Kg
    Advance per Micro-step1/16 Step (0.156?L)
    Max. Step Rate13867 (1/16 step) /sec
    Min. Step Rate16 (1/16) / 30 sec
    Max. Linear Rate130 mm/min
    Min. Linear Rate5?L/min
    Accuracy≤ 0.5% error in the condition of ≥ 30% of max. infusion distance
    Setting ModeRotary coded switch and membrane keypad
    Display128x64 graphic LCD
    Power SupplyAC100 -240V
    Operation ConditionTemperature: 5℃ - 40℃, Relative humidity: <80%
    Dimensions280×330×140 (mm)
    Weight5.3kg

    Other parameters

    TypeProduct CodeSyringeSyringe ID (mm)Flow Rate (μl/min-ml/min)
    LSP10-1B050346110μl0.500.001-0.0255
    25μl0.800.0025-0.0653
    50μl1.100.0048-0.1235
    100μl1.600.0101-0.2614
    250μl2.300.0208-0.5401
    500μl3.250.0415-1.0784
    1ml4.720.0875-2.2747
    2ml9.000.3181-8.2702
    5ml13.100.6739-17.522
    10ml16.601.0821-28.135